GEA NEO BREATH-II – Infant T-Piece Resuscitator Neopuff
Brand: GEA
Tipe: NEO BREATH-II
GEA NEO BREATH-II merupakan alat medis yang biasanya digunakan untuk memberikan bantuan oksigen untuk bayi.
Specification:
GENERAL PARAMETERS
– Intended users: Infants with a body mass of up to 10Kg
– Operating environment requirements: temperature: 18°C ~ 40°C, humidity: 5% ~ 95%
– Transport and storage environment requirements: temperature: -40°C ~ 60°C; humidity: up to 95%; atmospheric pressure 50 ~ 106kPa
– Protection against ingress of water: IPX4
– Total mass (including resuscitator and accessories): ≤ 6Kg
– Size (W x D x H): 290 x 180 x 370 mm
SYSTEM PARAMETER
– Gas supply: Medical oxygen and air (pipeline compressed gas supply system, or compressed gas cylinders)
– Gas supply input pressures range: 300 – 500kPa ( About 45 – 75Ps)
– Gas source flowrate: ≥ 50L/min
– Alarm: Single gas source fault alarm
– Low-pressure hose assemblies for use with medical use pressures range: 0-1000kPa
– Low-pressure hose assemblies for use with medical use flow range: 160-500L/min
AIR/OXYGEN MIXING FUNCTION
– Oxygen concentration setting range: 21% – 100%
– Accuracy: ≤ 3% VIV
– Reverse gas flow: Comply with the regulations of ISO11195:1995
– Flowrate setting range: 0 – 15/min, the level settings respectively is 0,5, 1, 2, 3, 4, 5, 6, 8, 10. 12,15 (L/min)
– Accuracy of florate output:
± 0,5U/min, @ 0,5, 1, 2, 3, 4 Umini
± 1U/min, @ 5, 6, 8. 10 L :
± 2L/min, © 12 and 15 L/min
VACUUM SUCTION FUNCTION
– Vacuum setting knob setting range: 0 – 18.67 1.33kPa (0 – 140 ± 10mmg)
– Free air flowrate: < 20L/min (at the maximum vacuum setting)
– Vacuum response time: When the input gas source pressure is 500kPa, vacuum in 10 seconds should be at least 17.34kPa (130mmHg)
– Scale range of vacuum gauge: 0 – 21kPa (0 – 160mmHg )
– Vacuum gauge accuracy: ± 5% of full-scale value
– Gas wastage: <28L/min (at the maximum vacuum setting)